Analyzing Samples 
26  Agilent Cary 630 FTIR Spectrometer User’s Guide 
Diamond ATR Sampling Accessory 
The Diamond attenuated total reflectance (ATR) sampling accessory 
takes advantage of the physical properties of light when encountering 
two materials with differences in index of refraction. When coming in 
contact with a sample having a different index of refraction, the 
infrared light creates an evanescent wave, which provides a very small 
and specific depth of penetration into the sample before reflecting 
back into the detector of the spectrometer. This small and consistent 
pathlength means no sample preparation is required to get good 
measurement results from a variety of samples. The key to obtaining 
good results with the Diamond ATR accessory is to make good contact 
between the sample and ATR crystal. The ATR technique can be used 
for analyzing liquids, pastes, powders and even some solid samples. 
The Diamond ATR accessory uses a type IIa diamond crystal as the 
interface between the sample and the infrared energy. Advantages of a 
diamond crystal are extreme hardness and chemical resistivity, plus it 
can accept samples with a pH range from 1 to 14. Therefore, hard or 
abrasive samples and even strong acids can be safely analyzed. The 
diamond in the Agilent ATR accessory is the world’s most durable 
substance and has been designed for maximum infrared signal. 
The Diamond ATR is a single-reflection ATR. It is most suitable for 
higher absorbing samples such as rubber, polymers, paints and fibers. 
Powder and solid samples are also best measured with the single-
reflection ATR, by using the sample press device, which applies high 
pressure to powder and solid samples to ensure good contact with the 
diamond crystal. The single-reflection ATR is also a good choice when 
the amount of available sample is limited. The single-reflection 
diamond has a 1 mm diameter sampling surface with 200 μm active 
area and provides approximately 2 μm depth of penetration for 
infrared energy at 1,700 cm
-1
. The single-reflection ATR protrudes 
slightly above the metal mounting plate. 
In the sample loading and cleaning mode, the sample press should be 
in the highest position, so the sample press tip is well above the 
diamond sampling window (see Figure 5). It is in this position that the 
sample mounting area can be easily accessed for loading the sample 
into place and for cleaning the sampling surfaces before the next 
sample is analyzed.