Analyzing Samples 
Agilent Cary 630 FTIR Spectrometer User’s Guide  53 
6  After completing the sample measurement, immediately clean the 
sample from the DialPath using the instructions provided 
previously. It is important to make sure both the top and bottom 
windows are free of any residue from the previous sample. 
For instructions on reviewing results and handling sample data, refer 
to the Agilent MicroLab PC Software User’s Guide. 
Diffuse Reflectance Sampling Accessory 
The Diffuse Reflectance sampling accessory provides quantitative and 
qualitative analysis of powdered or solid samples. This technique 
provides more sensitivity in a quantitative analysis than ATR, while 
still being easy to use.  
Powdered samples are commonly mixed in KBr powder at a level of 
approximately 3 to 5%. However, the loading of the sample in KBr can 
be adjusted to obtain the required sensitivity. The Diffuse Reflectance 
accessory is optimized for use with the Cary 630 FTIR optical system, 
providing the greatest sensitivity available for quick, accurate 
measurements. 
The Diffuse Reflectance accessory uses a four-position sample holder 
(see Figure 17), which slides into a slot on the right side of the 
accessory. The first position of the holder contains a diffuse 
reflectance gold mirror, which is used for background collection. The 
other three positions are used for filling with KBr/sample mixtures. 
Detents in the holder allow you to gauge which position is being 
measured. 
 
 
Figure 17. Diffuse Reflectance sample holder