Analyzing Samples 
Agilent Cary 630 FTIR Spectrometer User’s Guide  37 
In the case of a volatile sample, larger amounts of sample can be 
applied without concern of leakage or damage to the instrument. 
However, using the smallest amount of sample possible will ease 
in the cleaning process. 
CAUTION 
The Ge ATR accessory windows or metal mounting plate can be damaged by 
extreme samples. Only measure samples with a pH between 1 and 14. Do not 
leave extremely acidic or basic samples on the metal mounting plate for an 
extended time. 
 
CAUTION 
Although Ge is a hard material, the ATR window can crack under extreme 
pressure conditions. Ensure that the sample makes contact with the entire 
surface area of the crystal and not just a point within the crystal. Avoid using the 
sample press on samples that may be sharp or pointed. 
 
If the sample is a liquid or paste, the sample press does not need to be 
used at all. In this case, you can proceed with the analysis.  
If the sample is a powder or solid, the sample press must be in contact 
with the sample. To make contact, rotate the sample press knob in a 
clockwise direction until the pressure arm no longer goes down. A 
spring mechanism keeps pressure on the sample; take care not to over 
tighten the clamp. Once the pressure arm stops moving, apply no more 
than ¼ turn of the sample knob. 
4  For instructions on collecting a sample spectrum using the 
software, refer to the Agilent MicroLab PC Software User’s Guide. 
5  After completing the sample measurement, immediately clean the 
sample from the accessory using the instructions provided on 
Page 36. It is important to make sure both the Ge sampling 
window and the sample press tip are free of any residue from the 
previous sample. 
For instructions on reviewing results and handling sample data, refer 
to the Agilent MicroLab PC Software User’s Guide.