Chapter 2 173
Front-Panel Key Reference
Preset
User Preset This key is only available when the Preset Type key is set to User. Press
System, Power On/Preset, Preset Type, User.
User Preset loads the analyzer configuration that existed when Save
User Preset
was pressed. If Save User Preset has never been pressed, the
factory preset state is loaded. If the user preset state has been saved
but the load fails for any reason, the error message: Unable to load
user state is displayed in the status line and the state is reset to
whatever it was before the
Preset key was pressed. This can sometimes
happen if firmware has been upgraded or applications have been
(un)installed after the user preset state was saved.
Save User Preset can
be accessed by pressing
System, Power On/Preset.
Key Access:
Preset
Mode Preset
This key is only available when the Preset Type key is set to User. Press
System, Power On/Preset, Preset Type, User.
Mode Preset does not change the mode. It only resets the current mode
settings to the factory defaults and to continuous
sweeps/measurements.
Key Access:
Preset
Factory Preset
This key is only available when the Preset Type key is set to User. Press
System, Power On/Preset, Preset Type, User.
A factory preset is executed so all of the instrument modes are returned
to the factory default state (i.e. Spectrum Analysis Mode and
continuous sweep.) If you are not already in the spectrum analysis
mode, it switches to that mode. A factory preset will not reset
“persistent” functions such as GPIB address, time/date display style, or
auto alignment state to the factory defaults.
Key Access:
Preset
Save User Preset
This key is only available when the Preset Type key is set to User. Press
System, Power On/Preset, Preset Type, User.
This key saves the current state of the analyzer into the
User Preset
state for recalling when the instrument is user preset. After you save a
state here, you must go to the
Preset Type key and select User in order to
have this state used as the preset state.
Key Access:
Preset

Table of Contents

Other manuals for Agilent Technologies E4402B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4402B and is the answer not in the manual?

Agilent Technologies E4402B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4402B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals