EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B 113
SCPI Command Reference
:CALC{1-4}:TRAN:DIST:CENT
:CALC{1-4}:TRAN:DIST:CENT
Syntax :CALCulate{[1]|2|3|4}[:SELected]:TRANsform:DISTance:CENTer <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:TRANsform:DISTance:CENTer?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), sets the center value
used for the fault location display.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:TRAN:DIST:CENT 1E-8"
20 OUTPUT 717;":CALC1:TRAN:DIST:CENT?"
30 ENTER 717;A
Related commands :CALC{1-4}:TRAN:DIST:SPAN on page 119
:CALC{1-4}:TRAN:DIST:STAT on page 121
Equivalent key [Analysis] - Fault Location - Center
<numeric>
Description Center value
Range Varies depending on the frequency span, velocity factor, distance
unit, and the number of points.
Preset value 0
Unit Ft (feet) or m (meters)

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals