Appendix B 131
SCPI Command Reference
:CALC{1-4}:TRAN:TIME:KBES
:CALC{1-4}:TRAN:TIME:KBES
Syntax :CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:KBESsel <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:KBESsel?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), sets the shape of the
Kayser Bessel window using β used for the transformation function of the fault location
function.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:TRAN:TIME:KBES 3"
20 OUTPUT 717;":CALC1:TRAN:TIME:KBES?"
30 ENTER 717;A
Related commands :CALC{1-4}:TRAN:TIME:IMP:WIDT on page 130
:CALC{1-4}:TRAN:TIME:STEP:RTIM on page 137
:CALC{1-4}:TRAN:TIME:STAT on page 136
:CALC{1-4}:TRAN:DIST:KBES on page 116
Equivalent key [Analysis] - Fault Location - Window - Kaiser Beta
<numeric>
Description The value of β.
Range 0 to 13
Preset value 6