Appendix B 139
SCPI Command Reference
:CALC{1-4}:TRAN:TIME:STOP
:CALC{1-4}:TRAN:TIME:STOP
Syntax :CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:STOP <numeric>
:CALCulate{[1]|2|3|4}[:SELected]:TRANsform:TIME:STOP?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), sets the stop value used
for the transformation function of the fault location function.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:TRAN:TIME:STOP 2E-8"
20 OUTPUT 717;":CALC1:TRAN:TIME:STOP?"
30 ENTER 717;A
Related commands :CALC{1-4}:TRAN:TIME:STAR on page 135
:CALC{1-4}:TRAN:TIME:STAT on page 136
:CALC{1-4}:TRAN:DIST:STOP on page 124
Equivalent key [Analysis] - Fault Location - Stop
<numeric>
Description Stop value
Range Varies depending on the frequency span and the number of
points.
Preset value 1E-8
Unit s (second)