Appendix B 149
SCPI Command Reference
:CALC{1-4}:SRL:IMP:MAN
:CALC{1-4}:SRL:IMP:MAN
Syntax :CALCulate{[1]|2|3|4}:SRL:IMPedance:MANual <numeric>
:CALCulate{[1]|2|3|4}:SRL:IMPedance:MANual?
Description For the active trace of channel 1 (:CALC1) to channel 4 (:CALC4), sets the average cable
impedance to be used in the SRL calculation.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response {numeric}<newline><^END>
Example of use 10 OUTPUT 717;":CALC1:SRL:IMP:MAN 75"
20 OUTPUT 717;":CALC1:SRL:IMP:MAN?"
30 ENTER 717;A
Related commands :CALC{1-4}:SRL:IMP:AUTO on page 147
:CALC{1-4}:SRL:IMP:AUTO:CUT on page 148
Equivalent key [Analysis] - SRL - Manual Z
<numeric>
Description Averaging impedance value
Range 10 to 1000
Preset value 50 or 75
Unit Ohm