6-25
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
Init ignored;Sweep is already initiated.
Indicates that the list/sweep is currently initiated and sweeping, therefore the command
is not legal according to SCPI.
Init ignored;Sweep is already continuously initiated.
Indicates that the list/sweep is continuously initiated and sweeping, therefore the
command is not legal according to SCPI.
-221
Settings conflict;Cannot have uplink protocols while the control
frame is on. Frames 1-17 timeslots changed to downlink.
Indicates that, in TETRA mode, some of the timeslots are set to uplink. When the
control frame is turned on in TETRA, all timeslots in all frames must use downlink
protocols. The instrument enforces this by reconfiguring the timeslots that are not using
downlink protocols when the control frame is switched on.
Settings conflict;Cannot select uplink protocols while the
control frame is on. Selection changed to custom.
Indicates that in TETRA mode, an attempt has been made to select uplink protocols
while the control frame is switched on. TETRA requires that the control frame be
switched off in order to select uplink protocols.
Settings conflict;Continuous protocol timeslots cannot be turned
off.
Indicates that a continuous protocol timeslot is switched off while in TETRA mode.
When a continuous protocol has been selected for a timeslot in TETRA mode, all of the
timeslots must be switched on.
Settings conflict;Continuous & Discontinuous setting conflict.
All selected Continuous timeslots are now switched to
Discontinuous.
Indicates that while in TETRA mode, an attempt has been made to simultaneously
assign continuous and discontinuous protocols to the timeslots. Continuous and
discontinuous protocols cannot coexist in TETRA mode. If a discontinuous protocol is
selected for any timeslot, all timeslots’ protocols are changed to discontinuous
protocols.
Settings conflict;Discontinuous & Continuous setting conflict.
All selected Discontinuous timeslots are now switched to
Continuous.
Indicates that while in TETRA mode, an attempt has been made to simultaneously
assign discontinuous and continuous protocols to the timeslots. Discontinuous and

Table of Contents

Other manuals for Agilent Technologies ESG-D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESG-D Series and is the answer not in the manual?

Agilent Technologies ESG-D Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESG-D Series
CategoryPortable Generator
LanguageEnglish

Related product manuals