6-26
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
continuous protocols cannot coexist in TETRA mode. If a continuous protocol is
selected for any timeslot, all timeslots’ protocols are changed to continuous protocols.
Settings conflict;Frequency list and dwell list are of unequal
size. Set one list equal to size one, or make their sizes equal.
Indicates that the frequency list has more than one element and the dwell list has more
than one element, and they are not of equal size. If any of the frequency, power, or dwell
lists have more than one element, they must all have the same number of elements. A list
of a single element is the same as a list of equal size with the single element repeated the
necessary number of times.
Settings conflict;Frequency list and power list are of unequal
size. Turn one list off, set one to size one, or make their sizes
equal.
Indicates that the frequency list has more than one element and the power list has more
than one element, and they are not of equal size. If any of the frequency, power, or dwell
lists have more than one element, they must all have the same number of elements. A list
of a single element is the same as a list of equal size with the single element repeated the
necessary number of times.
Settings conflict;Power list and dwell list are of unequal size.
Set one to size one, or make their sizes equal.
Indicates that the dwell list has more than one element and the power list has more than
one element, and they are not of equal size. If any of the frequency, power, or dwell lists
have more than one element, they must all have the same number of elements. A list of a
single element is the same as a list of equal size with the single element repeated the
necessary number of times.
Settings conflict;The selected external trigger setting conflicts
with the previous setting.
Indicates that the external trigger has been set to positive edge for one trigger source and
negative edge for another trigger source.
Settings conflict;I/Q & AM WB cannot be on at the same time.
The modulation types I/Q and AM WB cannot be used at the same time. This error is
reported when the user attempts to activate one type of modulation while the other is
activated. The modulation type that was most recently requested will be turned on and
the modulation that was on at the time will be turned off.
Settings conflict;FM2/PM2 value set greater than FM1/PM1 value.
FM1/PM1 changed to match FM2/PM2 value.
The deviation of FM2/PM2 must always be less than or equal to the deviation settings