6-33
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
System error;Cannot change manual point until list mode error
condition cleared.
An error is keeping the sweep/list from being able to set the frequency and/or power.
Until the problem is addressed, the manual point cannot be changed.
System error;Unable to determine which attenuator is installed.
Indicates that an invalid attenuator identification code has been detected. Possible causes
include a loose attenuator control cable. The instrument will likely not produce the
proper output power levels. Report this error to the factory.
-311
Memory error;Unable to configure Save Recall registers from non-
volatile memory. Save Recall registers re-initialized.
Indicates that saved states are no longer usable. Delete explicitly using Catalog.
-315
Configuration memory lost;Persistent state preset. Using factory
defaults.
Indicates that the persistent state has been forced to return to factory preset values.
Configuration memory lost;Persistent state version is bad. Using
factory defaults.
Indicates that the persistent state version is not recognized as valid and is assumed to be
corrupt. The persistent state is reinitialized with the factory preset values.
Configuration memory lost;Persistent state checksum is bad. Using
factory defaults.
Indicates that the persistent state is corrupt and had to be reinitialized with the factory
preset values.
-321
Out of memory;Unable to verify instrument state file.
Indicates that an instrument state file could not be accessed and verified because of
insufficient memory. Reduce the size of any sweep lists and try again.
Out of memory;Memory catalog failed.
Indicates that there is not enough memory to complete a catalog listing. Reduce the size
of any sweep lists and try again.
Out of memory;Unable to display timeslot window.
Indicates that the instrument was unable to create part of the graphical user interface due
to an inability to allocate memory (possibly due to fragmentation). Please report the
circumstances to the factory. The instrument is still functional.

Table of Contents

Other manuals for Agilent Technologies ESG-D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESG-D Series and is the answer not in the manual?

Agilent Technologies ESG-D Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESG-D Series
CategoryPortable Generator
LanguageEnglish

Related product manuals