6-34
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
Out of memory;Unable to display protocol window.
Indicates that the instrument was unable to create part of the graphical user interface due
to an inability to allocate memory (possibly due to fragmentation). Please report the
circumstances to the factory. The instrument is still functional.
Out of memory;Unable to display format window.
Indicates that the instrument was unable to create part of the graphical user interface due
to an inability to allocate memory (possibly due to fragmentation). Please report the
circumstances to the factory. The instrument is still functional.
Out of memory;Cannot uncompress file.
Indicates that a STATE: file cannot be uncompressed because there is not enough
memory to run the decompression algorithm. Recall will fail and there will be no
instrument state change. Reduce the size of any sweep lists and try again.
Out of memory;Cannot precalculate frequencies. Try fewer
frequencies.
Indicates that memory was exhausted during frequency precalculation (used to speed the
process of sweep/list mode). List mode cannot run until either fewer frequencies have
been supplied or more memory becomes available and the same set of frequencies are
sent again,
FREQ:MODE CW is executed, or :FREQ:MODE LIST is executed.
Out of memory;Object Memory Area
Indicates that memory was exhausted during instrument power-on. Report the
circumstances to the factory.
Out of memory;List formation
The device was unable to allocate space for a lookup table, such as for list mode
precalculation. List mode cannot run until either fewer frequencies have been supplied
or more memory becomes available and the same set of frequencies are sent again,
FREQ:MODE CW is executed, or :FREQ:MODE LIST is executed.
Out of memory;PRBS xx/xx
There was not enough memory to apply a scramble to data for the baseband generator. In
this case, the scramble is not applied to the generator.
Out of memory;Display system out of memory. An abnormal display
may result. Memory consumption should be reduced.
There was not enough memory in the system to properly update the display. Some
inconsistencies may be seen. The size of any list/sweep should be reduced, and the
source should be preset to clear up any inconsistencies. Report the circumstances to the