6-35
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
nearest HP Sales and Service office.
Out of memory;Unable to check Data Generator memory.
There was not enough memory in the system to properly complete the data generator
memory test. This does NOT imply a data generator memory failure. Check all other
error messages to identify possible causes, discontinue list/sweep mode to free some
memory, and repeat the test.
-330
Self-test failed;Power supply self-test failure
Indicates that the self-test for a particular power supply voltage has failed. The
instrument is likely not functional. Contact the nearest HP Sales and Service office.
Self-test failed;EEPROM header checksum error <card_name>.
Indicates that the card identification header for a hardware card is incorrect. If the card is
not properly identified, the instrument is likely to be non-functional. Contact the nearest
HP Sales and Service office.
Self-test failed;Data Generator Memory Test @ 0x____
Indicates that the data generator memory failed. Modulation data produced by the data
generator may not be correct. However, if an
Unable to check Data Generator
Memory
error was also seen, this result is not conclusive. The address of the first
location that failed is reported. Contact the nearest HP Sales and Service office.
Self-test failed;Burst Generator Memory Test @ 0x____
Indicates that the burst generator memory failed. Modulation data produced by the burst
generator may not be correct. However, if an
Unable to check Burst Generator
Memory
error was also seen, this result is not conclusive. The address of the first
location that failed is reported. Contact the nearest HP Sales and Service office.
Self-test failed;Bad address position @ 0x____
Indicates that the data generator memory failed. Modulation data produced by the data
generator may not be correct. An address that appeared to have a failed address line was
reported. Contact the nearest HP Sales and Service office.
Self-test failed;Chips ___, ___ aliased @ 0x____
Indicates that the data generator memory failed. Modulation data produced by the data
generator may not be correct. An address that appeared to be aliased across multiple
memory chips has been reported. Contact the nearest HP Sales and Service office.
-430
Query DEADLOCKED
Indicates that a SCPI output queue has filled preventing further SCPI command