6-36
Error Messages
ESG-D Series Signal Generator Instrument-Specific Error Messages
execution, and there is no more room left in the corresponding SCPI input queue to
accept a query to read from the output queue. The system automatically discards output
to correct the deadlock.
201
Bad file number;Unable to check Data Generator memory.
Indicates that the instrument was not able to generate the pattern necessary to perform
the data generator memory test. This does NOT imply a data generator memory failure.
Report the problem to the factory.
208
I/O error;Unable to delete saved state from non-volatile memory.
No instrument state change.
Indicates that a STATe: file could not be deleted due to the file not being found, file
corruption, or another file-related problem. If the file is displayed by a memory catalog,
delete it explicitly.
I/O error;Save a state register ignored.
Indicates that a STATe: file could not be saved due to insufficient space, file corruption,
or another related problem.
I/O error;Delete empty sequence <sequence_name>. Delete sequence
ignored.
Indicates that the user has attempted to delete a sequence that is empty. This error
message is informational only. Typically, this error is reported several times when the
“Delete All Sequences” command is executed. If the file is displayed by Catalog, delete
explicitly.
I/O error;Delete a non-saved state register. Delete register
ignored.
Indicates that the user has attempted to delete an unused (empty) state. This error
message is informational only.
I/O error;Trailing zero found in <filename>. Fixing...
Indicates that a compressed state file has a zero at its end. This is a sign of file
corruption. The device fixes the problem by concealing the zero such that it no longer
triggers an error message. The file may be corrupt or unusable.
I/O error;Unable to recall from non-volatile memory. No
instrument state change.
Indicates that the state file is not readable and the recall was aborted.

Table of Contents

Other manuals for Agilent Technologies ESG-D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ESG-D Series and is the answer not in the manual?

Agilent Technologies ESG-D Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelESG-D Series
CategoryPortable Generator
LanguageEnglish

Related product manuals