EasyManuals Logo

Agilent Technologies G7121A User Manual

Agilent Technologies G7121A
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Agilent InfinityLab LC Series 1260 Infinity II FLD User Manual 49
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications (G7121A)
Table 5 Performance Specifications 1260 Infinity II Fluorescence Detector (G7121A)
Type Specification Comments
Detection type One signal wavelength (excitation and
emission)
Programmable single
wavelength (excitation
and emission)
fluorescence detector
Single wavelength
operation
RAMAN (H
2
O) > 500 (noise reference
measured at signal)
Ex=350 nm, Em=397 nm, dark value
450 nm, standard flow cell
RAMAN (H
2
O) > 3000 (noise
reference measured at dark value)
Ex=350 nm, Em=397 nm, dark value
450 nm, standard flow cell
Light source Xenon Flash Lamp, normal mode 20 W,
economy mode 5 W, lifetime 4000 h
Pulse frequency 296 Hz for single signal mode
74 Hz for economy mode
Maximum data rate 74 Hz
Excitation monochromator Range: settable 200 nm - 1200 nm and
zero-order
Bandwidth: 20 nm (fixed)
Emission monochromator Range: settable 200 nm - 1200 nm and
zero-order
Bandwidth: 20 nm (fixed)
Reference system in-line excitation measurement
Timetable programming Single signal wavelength, response time,
PMT Gain, baseline behavior (append,
free, zero)
Wavelength characteristic Repeatability +/- 0.2 nm
Accuracy +/- 3 nm setting

Table of Contents

Other manuals for Agilent Technologies G7121A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7121A and is the answer not in the manual?

Agilent Technologies G7121A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7121A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals