94 Agilent N5181A/82A MXG Signal Generators User’s Guide
Basic Digital Operation (Option 651/652/654)
Using Waveform Markers
Setting Marker Polarity
Setting a negative marker polarity inverts the marker signal.
1. In second Arb menu (page 87), press Marker Polarity.
2. For each marker, set the marker polarity as desired.
The default marker polarity is positive.
Each marker polarity is set independently.
See also, “Saving Marker Polarity and Routing Settings” on page 83.
As shown on page 93:
Positive Polarity: On marker points are high (3.3V).
Negative Polarity: On marker points are low (0V).
RF blanking always occurs on the low part of the signal regardless of the polarity setting.