3-26 Service Guide N5230-90014
Tests and Adjustments PNA Series Microwave Network Analyzers
System Verification N5230A
Interpreting the Verification Results
Figure 3-11 shows an example of typical verification results with Print Tabular Data
selected in the Printer Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of the device,
and the date tested.
Each S-parameter measurement result is printed with frequency tested, lower and upper
limit lines, the measured data, and the result of the test.
Figure 3-11 Example of Printed Tabular Verification Results