6.4 Menu Supervisor
Supervisor
Calibration
Season
Parameters
Maintenance
Language
Date format
6.4.1 Sub-menu Calibration
Measure the accuracy of the FLEXICOMPT AUTONOME+ during the calibration
with a gauge. It is available after a measurement. Data of the last measurement are
available.
NOTE: Only approved persons are permitted to remove the seal.
Calibration
Volume: xxx [Ech_Vol] Volume gauge: xxxx [Ech_Vol] Er: +xx.xx% Coeff: xx.xxxx p/L
Average flow: xxx [Ech_Deb]
Lo.flow vol.: xxx [Ech_Vol]
Mean temper.: xx.x °C
Nb.of pulses: xxxxx Pulses
Volume
: Display the volume;
Gauge volume
: Enter the volume read on the
calibration mean;
Er
: Display the error in %;
Coeff
: Coefficient to be set only by an
authorized person in METROLOGICAL mode, if required
Average flow
: Average flow of the measurement
Mean temper
: Mean temperature of the measurement