FORM DOC 011 A
MU 7033 EN M
FLEXICOMPT AUTONOME+
Page 26/42
This document is available on www.alma-alma.fr
7.3.4 Sub-menu Direction
Do not enable this feature.
Direction
Enable
OFF
ON
Enable
: Choose
OFF
7.3.5 Sub-menu Temperature
This menu is an option. It is used to calibrate the temperature into the FLEXICOMPT
AUTONOME+. See maintenance sheet FM 8513
Temperature
Calibration
Enable
Ref. tempera.
+15.0 °C
Vol. disp.
OFF
ON
Vm
Vb / V15 / V20
+00.00 °C
Enable
: Enable or disable the product temperature control
Calibration
: The temperature calibration can be done either on two measuring points
or on a single measuring point.
Calibration on two temperature measuring points:
The measure must be done outside the range -20 to +50°C. First point at t<-
20°C, second point at t>+50°C.
Calibration on a single temperature measuring point:
The measure must be done in the range -20 to +50°C.
Ref. tempera
: Reference temperature (°C)
Vol. disp.
: Choose the volume displayed in USER mode:
Vm
: Volume in metering conditions
Vb
: Volume converted to the reference temperature