FORM DOC 011 A
MU 7033 EN M
FLEXICOMPT AUTONOME+
Page 20/42
This document is available on www.alma-alma.fr
Display LCD display fault
DISPLAY MEANING ACTION
High flowrate
Low flow high
Gas
Dry metering
Coil
Temperature
Flowrate greater than the setting maximum flowrate
Flow greater than 20m
3
/h while GDh dry
GDh is wet but GDl is dry
When using a pump. The volume of gas is greater
than the minimum measured quantity
Loss of pulse transmitter signal
Temperature less than -20°C or greater than 50°C
Check the setup
Check the hydraulic configuration / Check the detector status
Stop metering
Flowrates
Frequency
Flow setting fault
Frequency fault
Coefficients Difference two coefficients is greater than 0.5% Check the coefficients setup
Check the parameters
Sensor 2
Low gas detector fault (GDl)
Low flowrate
Sensor 1
Overflow
Flow rate less than the setting minimal flow rate
High gas detector fault (GDh)
Volume greater than 4 194 304 liters
Check the hydraulic configuration and the flowing
Use the maintenance menu to check the status of the detector
Reset the device
Use the maintenance menu to check the status of the detector
USER
Check the parameters
Check the setup
If steady alarm, substitution of the UNI-2
Check the temperature sensor (measure and calibration)
Check the connection with the pulse transmitter
REPARATOR
Date time Loss of date and time
Set date and time in metrological mode or use the menu
‘Connect>Sart GPS’ to switch on the GPS.
done outdoors. It lasts one minute to synchronize the clock
Metering Problem of metering with the meter Check the setup
Failure Problem with the transfer of the files to the CTD+ See GU 7110
Watchdog Fault with card If steady alarm, substitution of the UNI-2
Program Error on the cheksum of the metrological data If steady alarm, substitution of the UNI-2
RAM Saved memory fault If steady alarm, substitution of the UNI-2
Memory Bad writting into the memory If steady alarm, substitution of the UNI-2
Low Battery The battery is no more charging
Metrological Loss of configuration
Substitution of the battery
If steady alarm, substitution of the UNI-2
Totalizer Totalizer fault If steady alarm, substitution of the UNI-2
Memory Default
Problem with the measurement integrity: loss of
backup data concerning the last measurement
If steady alarm, substitution of the UNI-2
Micro SD card Problem with the micro SD card Check the micro SD card is in. Try another one if necessary
Init Bluetooth
Bluetooth module initialization problem or low
batteries
Charge the batteries or restart the UNI-2 via the menu
Supervisor>Maintenance>Reboot
Init GPS GPS module initialization problem or low batteries
Charge the batteries or restart the UNI-2 via the menu
Supervisor>Maintenance>Reboot
Init Wi-Fi Wi-Fi module initialization problem or low batteries
Charge the batteries or restart the UNI-2 via the menu
Supervisor>Maintenance>Reboot