CHAPTER 2: INSTALLATION AND SETUP
UNIFLOW
®
4 PULSED HEAT CONTROL
2-6 990-908
Input Power Connections
CAUTION
Be sure that the source power is appropriate for your Control model.
Attach an electrical plug to the power cable supplied in the Ship Kit. Connect the Control power cable
to a single-phase, 50/60Hz power source. The nominal voltage range for each model is set at the
factory: either 100 to 120 VAC, or 200 to 240 VAC.
Connect the Transformer (Uniflow 4R Version Only)
Before you start, be aware that electrical noise may cause error messages such as the MAX TEMP
alarm, and the temperature control will not be as good. To prevent electrical noise, follow the
instructions for proper cable placement in the following instructions.
The 24 VDC power cable must be
connected to the 24 VDC connectors on
the back of the transformer and the back
of the control.