EasyManua.ls Logo

Ametek 888

Ametek 888
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-2 | Model 888 ATEX IECEx Tail Gas Analyzer
Operational Alarms
Op[erational alarms indicate a condition that prevents normal analyzer
operation. The condition may or may not require user intervention to
clear the problem. Operational alarms include:
Low Light Intensity on a detector channel
Sample Cell Temperature is outside the minimum/maximum values
Demister Cell Temperature is outside the minimum/maximum values
Sample Flow is outside the minimum/maximum values
Excessive Sample Cell Pressure indicating a possible probe blockage
Manifold Pressure is too low: pneumatic valves may not be operating
properly.
Level 2 Alarms
A Level 2 alarm condition indicates a serious problem with the analyzer.
The analyzer responds by opening both the system alarm and data valid
alarms. The state machine, however, does not jump to the Not Ready
state. A Level 2 alarm condition can occur when there is a failed zero or
span calibration cycle or if a service technician stops the period timer.
Level 2 alarms include:
Period Timer stopped
Span calibration factor is outside the minimum/maximum values
Zero calibration failure
Level 3 Alarms
A Level 3 alarm condition is a warning that a more serious problem is im-
minent. There is a visual warning on both the local and remote displays.
The operator should take action to prevent the problem from becoming a
Level 1 or Level 2 alarm. Level 3 alarms include:
Excessive Flash Warning indicating that the light intensity is dropping
Low Flange Temperature warning indicating the possibility of a blockage
Detector Temperature alarm indicating that the accuracy of the con-
centration measurements may be affected.

Table of Contents

Related product manuals