Specications
| 2-3
Utility
Requirements
Power Consumption (specied at time of purchase)
120 VAC (105-132 VAC), 47-63Hz, 500W
OR
240 VAC (209-264 VAC), 47-63Hz, 500W
Instrument Air
Supply Instrument Air 60 lb max. (60psig) (414kPa)
Pressure 379-517 kPa (55 to 75 psig / 3.8-5.2 barg)
Air Consumption (Fast Purge) = 1.023 kg/min and
2.262 lbm/min
Air Consumption (Trickle Purge) = 0.00138 kg/min
and 0.00304 lbm/min
Air Consumption (Zero and Aspirator at 75 psig) =
0.169 kg/min and 0.373 lbm/min
Air Quality: As per ANSI/ISA standard 7.0.01-1996)
Steam
Ball Valve Jacket Pressure: 517-690 kPa (75 to 100
psig \ 5.2-6.9 barg)
Air Consumption (Fast Purge) = 1.023 kg/min and
2.262 lbm/min
Air Consumption (Trickle Purge) = 0.00138 kg/min
and 0.00304 lbm/min
Air Consumption (Zero and Aspirator at 75 psig) =
0.169 kg/min and 0.373 lbm/min
Blowback Pressure: 207-345 kPa (30-50 psig / 2.1-
3.5 barg).