AMETEK CTS Quick start guide - Conducted
1.00 10 / 50
The position of the voltage selector must correspond with the mains.
If you change the mains voltage, replace the fuses according the
recommended value on the number plate.
First - check the 115 V / 230 V voltage selector where applicable!
Do not connect 230 VAC into the 115 VAC units.The result could be a
severely damaged unit.
Figure 4 - Power Settings
If power to the equipment of Class I is supplied by an auto-transformer which is connected to a higher supply voltage, the base
of the auto-transformer must be connected to the neutral of the power supply.
Equipment of Class I, which is supplied with a 3-phase power cord, must only be connected to a grounded receptacle. The
protective ground connection must not be disconnected or otherwise interrupted.
Each interruption or disconnection of the protective ground connection inside or outside the equipment may cause electric
shock.
Connecting different units with each other or to a personal computer for remote control may be performed only with cables
recommended and/or supplied by the manufacturer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek EM TEST compact NX and is the answer not in the manual?

Ametek EM TEST compact NX Specifications

General IconGeneral
BrandAmetek
ModelEM TEST compact NX
CategoryPortable Generator
LanguageEnglish

Related product manuals