Portable Thermometers
User Guide
Cyclops L
Appendix 1 - 2
(ii)PacketIdentier1char
‘C’Classic–whilekeypressed
‘H’Hold–whenkeyreleased
(iii)Units1char
‘C’°Celsius
‘F’°Fahrenheit
‘R’Rankine
‘K’Kelvin
(iv)Mode1char
‘I’ Instantaneous
‘P’ Peak
‘V’ Valley
A’ Average
‘S’ StatisticalProcessingFunction
(v)Data1degresolutionCyclops6chars
:tenthsto999.0,degreesabove1000‘+100.0’‘+
1500
0.1degresolutionCyclops7chars
 :intenths‘+100.3’‘+1500.8’
Over-range: +/-¯¯¯i.e.0xAF,SP,0xAF,SP,0xAF
Under-range:+/-___ i.e.0x5F,SP,0x5F,SP,0x5F
 (vi)Termination2chars0x0D,0x0A
Thereceivingdeviceshouldsenda‘*‘handshaketoindicate‘H’hold–i.e.key
releasedanddatalogged
Advanced – Burst Mode packets
Whentriggerispressedinburstmode,acontinuousstreamoflinesofdata,
every30msisoutput
1degresolutionexample
‘B’ ‘3’ ‘C’ ‘I’ ‘+972.0’ ’CR’’LF’
0.1degresolutionexample
‘B’ ‘3’ ‘C’ ‘I’ ‘+1972.3’’CR’’LF’
Characterinterpretationasaboveexcept
(i) Packet ‘3’-advanced–burstmodepacket
Whentriggerisreleasedinburstmode,a2lineterminationpacketisoutput