Portable Thermometers
Cyclops L
ERROR CODES
A2
Error Codes
Atswitch-on,theCyclops instrumentwillcheckallstoreddatainitsnon-
volatilememoryandanydetectederrorswillbeshownnearthebottomofthe
sidedisplayscreen.
Error Code 1
Description: UnknownCyclopsType
Ifthiscodeisdisplayedtheinstrumentwillbeunabletoperformanyfurther
anyfunction.
Iftheerrorpersists,returntheCyclopstoAMETEKLandforrepair.
Error Code 2
Description: LostCalibrationData
The Cyclopswillcontinuetowork,butwiththedisplayinverted(i.e.white
onblack).However,thecalibrationoftheinstrumentwillbeincorrectand
readingswillonlybeapproximate.Itwillnotmeetthetemperatureaccuracy
specications.Ifthiserrorpersists,returntheCyclopstoAMETEKLandfor
repair.
Error Code 3
Description: LostUserSetting
The Cyclopsnormallystoresusersettings(accessedviathemenu)innon-
volatilememory.Iftheseareinconsistent,theusersettingswillrevertto
factorydefaults.Theoperatorcanre-enterthedesiredusersettingstoallow
the Cyclops toworkasintended.