EasyManua.ls Logo

Ametek Land Cyclops L Series

Ametek Land Cyclops L Series
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Portable Thermometers
Cyclops L
ERROR CODES
A2
Error Codes
Atswitch-on,theCyclops instrumentwillcheckallstoreddatainitsnon-
volatilememoryandanydetectederrorswillbeshownnearthebottomofthe
sidedisplayscreen.
Error Code 1
Description: UnknownCyclopsType
Ifthiscodeisdisplayedtheinstrumentwillbeunabletoperformanyfurther
anyfunction.
Iftheerrorpersists,returntheCyclopstoAMETEKLandforrepair.
Error Code 2
Description: LostCalibrationData
The Cyclopswillcontinuetowork,butwiththedisplayinverted(i.e.white
onblack).However,thecalibrationoftheinstrumentwillbeincorrectand
readingswillonlybeapproximate.Itwillnotmeetthetemperatureaccuracy
specications.Ifthiserrorpersists,returntheCyclopstoAMETEKLandfor
repair.
Error Code 3
Description: LostUserSetting
The Cyclopsnormallystoresusersettings(accessedviathemenu)innon-
volatilememory.Iftheseareinconsistent,theusersettingswillrevertto
factorydefaults.Theoperatorcanre-enterthedesiredusersettingstoallow
the Cyclops toworkasintended.
User GuideAppendix 2 - 0

Table of Contents

Related product manuals