EasyManua.ls Logo

Ametek PARSTAT MC - 6 Available Options; FRA, Available with PMC-1000; Auxiliary Cables Options; Advanced Auxiliary Interface Option

Ametek PARSTAT MC
62 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PARSTAT MC Hardware Manual
PARSTAT MC Hardware Manual 60
6. AVAILABLE OPTIONS
The factory installed options available for the PMC-1000 and PMC-1000/DC are as follows:
1.) FRA for Electrochemical Impedance Spectroscopy, with the PMC-1000.
2.) Auxiliary Cables
3.) Advanced Auxiliary Interface
4.) VersaSTAT LC Low Current Interface
6.1. FRA, available with PMC-1000
The FRA Option provides the hardware necessary to perform the ac-based techniques of Potentiostatic
Electrochemical Impedance Spectroscopy (PEIS) and Galvanostatic Electrochemical Impedance
Spectroscopy (GEIS). This function is part of the PMC-1000.
6.2. Auxiliary Cables Options
There are 2 different 9-pin connections to each PMC-1000 and PMC-1000/DC potentiostat channel.
The PMC AUX01 is a 1-meter cable that splits the DB9 connection of the Digital Ins/Outs into convenient
separate BNC connections.
The PMC ALG01 is a 1-meter cable that splits the DB9 connections of the Analog Ins/Outs into convenient
separate BNC connections.
Consult Chapter 5 for specific pin outs and function of the different connections.
The cell cable is provided standard with the PMC-1000.
6.3. Advanced Auxiliary Interface Option
The Advanced Auxiliary Interface (AAI) Option allows for an additional four (4) A/D inputs for the
VersaStudio software to acquire along with data recorded directly from the PMC-1000 and PMC-1000/DC
channels.
The AAI is an external unit that plugs directly into the same PC as that interfaced by the PMC-1000 into an
available USB port. The VersaStudio auto-detects the AAI, and collects data on the additional four A/D
inputs along with data directly from the PMC hardware (E, I, AUX, etc.).
The AAI is a +/-10V DC input, 12-bit unit (+/-5mV resolution) system, and is recommended for when several
variables need to be collected concurrently with the electrochemical data, such a pH, temperature, etc.
6.4. VersaSTAT LC Low Current Interface
The VersaSTAT LC Low Current Interface is a plug-and-play add-on option that provides current ranges
down to a minimum range of 4 pA when used with the VersaSTAT MC. Please see product brochure for
additional details.