Remote Operation
M370430-01 Rev B 5-67
5
:PROG:STAT RUN
The output after the last command might be as follows:
AMETEK, XG (or XTR) 150-5.6, SN# E00123456, 1.00 Build
10, 21/ 11/2005
AMETEK, XG (or XTR) 150-5.6, SN# E00123456, 1.00 Build
10, 21/ 11/2005
Important: Execution of the program may be terminated at any time by
pressing the Esc key in the MS Windows Hyper Terminal window. The power
supply can be controlled during program execution.
Deleting an Auto Sequence Program:
It is possible to delete the current Auto Sequence Program using the
SCPI noted below.
Command:
[:]PROGram[<channel>]:DELete[:ALL]
Repeating an Auto Sequence Program:
An Auto Sequence Program can be configured to repeat any
number of times or infinitely. The default at power up is to have the
repeat count default to once. The program will repeat if the repeat
count can be changed using the : PROGram [<channel>] :
REPeat command. If a program is running or pause the : PROGram
[<channel>] : REPeat command will not work.
Commands:
[:]PROGram[<channel>]:REPeat {?|<count>|INFinity}
Where:
<count> is the repeated count from 1 to 65 534.
INFinity is the character mnemonic for repeating without end.
For Example:
:PROG:STAR
*IDN?
*IDN?
:PROG:STOP