Calibration and Troubleshooting
7-26 M370430-01 Rev B
Restore Factory Calibration
To restore factory calibration, use the SCPI command for returning the
power supply to factory calibration settings.
The SCPI Command (s) for these instructions are:
[:]CALibration:RESTore
Restore Default Calibration
To restore default calibration (unit gain and zero offset for all
calibration parameters), use the SCPI command for returning the power
supply to default calibration settings. This can be useful when
recalibrating the unit to a specific load but causes ALL calibration
values to return to default values. All types of calibration must be done
to restore the unit to fully functional condition. It is strongly
recommended that the calibration coefficients be saved before using this
command. (See “Storing and Loading Calibration Parameters” on page
7-25).
The SCPI Command (s) for these instructions are:
[:]CALibration:DEFault
User Diagnostics
If your power supply is not performing as described in this Manual, read
through the procedures and checks in this chapter before calling your
service technician. These procedures are limited to operator level
functions only. They do not require cover-off servicing of the unit.