A-16
M370430-01 Rev B
SCPI Command Reference
Table A-1 IEEE 488.2 Commands (Continued)
Trigger
*TRG
Trigger system
N/A
Self-Test
Query
*TST? [:]SYSTem[<channel>]:TEST?
Perform unit Self Test
(Pass=0).
Yes
Wa i t To
Continue
*WAI [:]SYSTem[<channel>]:WAIT
Wait to continue command.
N/A
Table A-2 Readback Commands
Function
SCPI Command
Description
Parameter and
Range
Query
Read Output
Current
[:]MEASure[<channel>][:SCALar]:CURRent
[:DC]?
Measure and readback
output current at the
output terminals.
Yes
Read Output
Voltage
[:]MEASure[<channel>][:SCALar][:VOLTage]
[:DC]?
Measure and readback
output voltage at the
output terminals.
Yes
Read APG
Input Voltage
[:]MEASurement
[:SCALar]:APRogram
[:VOLTage][:DC]?
Measure and readback the
voltage being input to the
non isolated voltage
analog programming
input pin.
Yes
Read Isolated
APG Input Vo
lt a g e
[:]MEASurement [:SCALar]:APRogram
[:VOLTage]:ISOLated [:DC]?
Measure and readback the
voltage being input to the
isolated voltage analog
programming input pin.
Yes
Read APG
Input Voltage
[:]MEASurement
[:SCALar]:APRogram
:CURRent[:DC]?
Measure and readback the
voltage being input to the
non isolated current
analog programming
input pin.
Yes
Read APG
Input Voltage
[:]MEASurement
[:SCALar]:APRogram :CURRent:ISOLated
[:DC]?
Measure and readback the
voltage being input to the
isolated current analog
programming input pin.
Yes