Serial Communications
| A-9
Config_flags
S2
System configuration flags (1=true,
0=false)
bit 0 – auto verify cycle time is
programmed in hours, 0 if in days
bit 1 – Auto cal option is installed
bit 2 – Combust module is installed
bit 3 – Future use
bit 4 – Future use
bit 5 – RTD type combustible detector, 0=
Hot wire detector
bit 6 – WDG type sensor, 0 is HP type
sensor
bit 7 – combustibles are measure in
percent,0 = combustibles are measured in
PPM
bit 8 – relays energize on alarm. If not,
they deenergize.
bit 9 – auto calibration time is set, an
auto calibration will happen at the set
time if the auto calibration is enabled
bit 10 – auto calibration timer is enabled,
since this is enabled, an auto calibration
will occur at the auto calibration time.
(0=true, 1=false)
bit 11 – real time clock is set.
bit 12 – Step Function enabled (only for
RSS)
bit 13 – auto verify time is set, an auto
verify will happen at the set time if the
auto verify is enabled.
bit 14 – auto verify timer is enabled, if
this is enabled an auto verify will occur
at the auto verify time. (0=true, 1=false)
bit 15 – auto calibration cycle is
programmed in hours, 0 if in days
R/W