Serial Communications
| A-15
29/41 Recov_time Amount of time for recovery after a
calibration in seconds
IQ
Amount of time for recovery after a
calibration in MMSS
S2
R/W
2A/42 O2_gas1 Oxygen span gas setpoint value, always in
percent
R/W
2B/43 O2_gas2 Oxygen zero gas setpoint value, always in
percent
R/W
2C/44 Co_gas1 Combustibles PPM span gas setpoint value
S2
Combustibles span gas setpoint
IQ
R/W
2D/45 Co_percent_gas Combustibles percent span gas setpoint
value
S2
R/W
2E/46 Me_percent_gas Methane percent span gas setpoint value R/W
2F/47 Cal_cgas1 Oxygen span gas setpoint for last
calibration
Read only
30/48 Cal_span Oxygen span gas reading for last
calibration
Read only
31/49 Ver_cgas1 Oxygen span gas setpoint for last verify Read only
32/50 Ver_span Oxygen span gas reading from last verify Read only
33/51 O2_span Oxygen span gas millivolts during last
calibration
Read only
34/52 Cal_cgas2 Oxygen zero gas setpoint value for last
calibration
Read only
35/53 Cal_zero Oxygen zero gas reading from last
calibration
Read only
36/54 Ver_cgas2 Oxygen setpoint value used during last
verify
Read only
37/55 Ver_zero Oxygen zero gas read during last verify Read only
38/56 O2_zero Zero gas millivolts read during last
calibration
Read only
39/57 Cal_cgas3 Combustibles span gas setpoint during last
calibration
Read only
3A/58 Cmbspan Combustibles span value during last
calibration
Read only
3B/59 Ver_cgas3 Combustibles span gas setpoint for last
verify
Read only
3C/60 Vcmbspan Combustibles span gas read during last
verify
Read only
3D/61 Cmbzero Combustibles zero gas read during last
calibration
Read only
3E/62 Vcmbzero Combustibles zero gas read during last
verify
Read only
3F/63 Cal_cgas4 Methane span gas setpoint during last
calibration
Read only
40/64 Mthspan Methane span value read during last
calibration
Read only
41/65 Ver_cgas4 Methane span gas setpoint for last verify Read only
42/66 Vmthspan Methane span value during last verify Read only
43/67 Mthzero Methane zero value during last calibration Read only
44/68 Vmthzero Methane zero value during last verify Read only

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox Series 2000 WDG-IVC and is the answer not in the manual?

Ametek Thermox Series 2000 WDG-IVC Specifications

General IconGeneral
BrandAmetek
ModelThermox Series 2000 WDG-IVC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals