EasyManuals Logo

Ametek Thermox Series 2000 WDG-IVC User Manual

Ametek Thermox Series 2000 WDG-IVC
193 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #175 background imageLoading...
Page #175 background image
A-20 | Thermox Series 2000 WDG-IV /I VC / IVCM
62/98 Cal_variable The calculated percent value that 0 mv
would correspond to with the cell that is
currently being used. This value can vary
between different cells. It is calculated
during calibration and used in the oxygen
calculations.
Read only
63/99 Option_flags
Option flags (1=true, 0=false)
bit 0 – combustible detector installed
bit 1 – Data module installed
IQ
Dual current output installed
S2
bit 2 – 2 span value RCU
IQ
auto cal option installed
S2
bit 3 – Methane detector installed
IQ
3-in-1 option installed
S2
bit 4 – Dig – ma module installed
IQ
bit 5 – HART module installed
IQ
bit 6 – RS485 module installed
IQ
bit 7 – 3 span value RCU
IQ
Invalid option module
S2
Read only
64/100 Tspan_mv Cell mv value measured with the span gas
flowing during last calibration.
Read only
65/101 Tzero_mv Cell mv value measured with the zero gas
flowing during calibration.
R/W
66/102 Mthvalue Read only
67/103 M_store Read only
68/104 Cal_temp Last calib temperature R/W
69/105 CJTemp Cold Junction temperature Read only
6A/106 Box_temp Box Temperature Read only
6B/107 Cmb_Kfact Combustible K Factor R/W
6C/108 Mth_Kfact Methane K Factor
IQ
R/W
6D/109

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox Series 2000 WDG-IVC and is the answer not in the manual?

Ametek Thermox Series 2000 WDG-IVC Specifications

General IconGeneral
BrandAmetek
ModelThermox Series 2000 WDG-IVC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals