EasyManua.ls Logo

Ametek Thermox Series 2000 WDG-IVC - Manual Calibration and Aspirator Air Connections

Ametek Thermox Series 2000 WDG-IVC
193 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation | 3-15
Manual calibration and aspirator air connections
If you have a remote calibration unit (RCU), skip the remainder of
this section and proceed to the “RCU Installation Instructions” ap-
pendix.
If performing manual calibrations, do the following:
1. Connect aspirator air to the aspirator air inlet on the sensor (see Fig-
ure 3-9). Adjust the aspirator air regulator to the minimum pressure
required to get a response from the analyzer; typically this is 5 psi. Do
not turn on the aspirator until the sensor has been turned on and is
hot -preferably 24 hours.
2. A calibration gas inlet port is provided to allow you to calibrate the
system as shown in Figure 3-9. During calibration, you turn off the
aspirator air supply and inject your calibration gases into the calibra-
tion gas inlet port on the sensor. During normal operations, you plug
this calibration gas inlet port.
• You must provide a protective gas supply with an alarm for a loss of
pressure indication in order to meet NFPA 496 requirements.
• Do not exceed the maximum temperature rating listed on the warning
label on the Z-purge enclosure
NOTE

Table of Contents

Related product manuals