EasyManuals Logo

Ametek Thermox TM2000 User Manual

Ametek Thermox TM2000
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #167 background imageLoading...
Page #167 background image
Using Series 2000 RCU Signals | D-1
USING SERIES 2000 RCU SIGNALS
You can use Series 2000 control unit signals to control your own switching
valve to switch from the sample to calibration gases. Note, however, that
we recommend that you use the Factory-Provided Remote Calibration
Unit (RCU) as the best way to control the calibration process.
You must have and auto calibration card installed in your Series 2000
controller to be able to use these signals.
RCU signals allow you to control the valves you use to switch between
the process gas and the calibration gases. See the Controller/User Interface
chapter for help on how to set up the calibration process from software.
Figure D-1.
Control unit signals
to initiate a remote
calibration.
If measuring PPM gases, use a packless valve to switch between the
sample gas and the calibration gas.
NOTE
Control Signals *High or Low Signal from Control Unit
Calibration Control Signal
(labeled Aspirator on control unit)
System Not in Cal. - Low
System Calibration - High
Zero (Low) Cal Gas
(labeled Zero on control unit)
Calibrating Zero Gas - High
All other Times - Low
Span (High) Cal Gas
(labeled Span on control unit)
Calibrating Span Gas - High
All other times - Low
*High signal is 15 volts (sources current), low signal is 0 volts (no current sourced).
Maximum source curent - 250 mA.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox TM2000 and is the answer not in the manual?

Ametek Thermox TM2000 Specifications

General IconGeneral
BrandAmetek
ModelThermox TM2000
CategoryAnalytical Instruments
LanguageEnglish