5-4 | Thermox TM2000 Oxygen Analyzer
• Temporarilyenableordisableautocalibrationsorverifications(SetCal
Timers/Auto Cal On/Off).
• Definehowlongcalibrationgasesshouldbeappliedtothesensordur-
ing an auto calibration or verification (Set Cal Timers/Cal Gas Dura-
tion).
• Definerecoverydurationtimes(SetCalTimers/RecoveryDuration).
• Defineverifygasvaluestomatchknownvaluesofverifygascylin-
ders; enter acceptable verify range (Configure Verify).
Aborting a Calibration
If you press the Calibrate key while the system is performing a calibration,
you will be have the option to either abort or to continue the calibration.
All other menu functions are temporarily disabled during a calibration