EasyManuals Logo

Ametek Thermox TM2000 User Manual

Ametek Thermox TM2000
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
5-20 | Thermox TM2000 Oxygen Analyzer
Set Cal Timers
The Set Cal Timers submenu from the Calibrate key allows you to do the following:
• Setautotimerfordesiredautocalibration/verificationcycletimes(Set
Auto Timers).
• Setthecurrenttimeanddate(SetTimeandDate).
• Enableordisabletheautocalibrationcycle(AutoCalOn/Off).
• Definehowlongcalgasesshouldflowthroughthesensorduring
automatic calibrations or verifications (Cal Gas Duration).
• Definearecoveryduration,duringwhichtimethesensorswitches
from measuring calibration gas readings to measuring process read-
ings without affecting the analog output or triggering of alarms (Re-
covery Duration).
Set Cal
Timers
Set Auto
Timers
Set Time
& Date
Auto Cal
On/Off
Recovery
Duration
Cal Gas
Duration
Figure 5-3.
Set Cal Timers menu
options

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek Thermox TM2000 and is the answer not in the manual?

Ametek Thermox TM2000 Specifications

General IconGeneral
BrandAmetek
ModelThermox TM2000
CategoryAnalytical Instruments
LanguageEnglish