6-2 | Thermox TM2000 Oxygen Analyzer
System Messages
Auto Cal Pending
This message indicates that an automatic calibration is pending. Once the
current calibration or verification is complete, the automatic calibration
will begin. This message will only appear if the Auto Calibration option is
installed and enabled.
Auto Verify Pending
This message indicates that an automatic verification is pending. Once the
current calibration or verification is complete, the automatic verification
will begin. This message will only appear if the Auto Calibrate option is
installed and enabled.
Calibration Aborted
This message indicates that the calibration has been aborted by the user.
Calibration Required
This message is displayed to inform the operator that a regular calibra-
tion is required (after a primary calibration is performed you must al-
ways perform a regular calibration). During a primary calibration the cell
temperature is changed and you must wait for one hour before a system
calibration can be performed. The system calibration will determine the
constants needed for the Nernst equation.
If you see this message, select the Calibrate Key and perform a calibration.
This message will not clear from the display by removing or restoring
power to the control unit.
Cal Verify Aborted
This message notifies the operator that the verification was successfully
aborted by the system.