Specications | 2-1



4 line x 20 character vacuum fluorescent. Displays combinations of oxygen
(0.1 PPM O
2
to 100%, auto-ranging), time and date, cell temperature, user
programmable text, thermocouple mV or cell mV. Password protection,
programmable pressure compensation and context-sensitive help are also
provided.

Two isolated linear current outputs. Select O
2
, cell temperature, thermo-
couple mV or cell mV. Each output can be 4-20 mA, 0-20 mA, 20-4 mA or
20-0 mA, and is fully scalable. Hold or track during calibration and select
degree of damping. Maximum load 1200 ohms.

Two independent oxygen alarms, each high or low selectable. One alarm
can be assigned as oxygen, calibrate or verify. Set relays to energize or de-
energize on alarm.

0.5A, 30V, 10VA max. noninductive load, AC or DC