SECTION 6 PERFORMANCE TESTS
6-30
(4) Procedure
LOG display linearity
Step Procedure
1 Press the [Preset] key.
2 Operate All Cal.
3 Set the MG3633A to 100 MHz and 0 dBm.
4 Set the MN510C to 0 dB.
5 Set the spectrum analyzer as shown below:
Center Freq............................................100 MHz
Span.......................................................... 10 kHz
Reference Level ........................................ 0 dBm
Attenuator................................................... 10 dB
RBW........................................................... 3 kHz
VBW ........................................................ 300 Hz
6 Press the [ CF] key to set the spectrum waveform peak to the center of the screen.
7 Adjust the MG3633A output level so that the marker level reading is 0.0 dBm.
8 Press the [Marker] key sequentially to set the marker to marker after the sweep is com-
pleted.