SECTION 6 PERFORMANCE TESTS
6-55
REF LEVEL
ATT
Attenuator
setting
Spectrum analyzer setting
Correction
factor of
attenuator
Marker
readout
Error Deviation
–10 dBm 50 dB 0 dB dB dBm dB dB
–20 dBm 40 dB 10 dB dB dBm dB dB
–30 dBm 30 dB 20 dB dB dBm dB dB
–40 dBm 20 dB 30 dB dB dBm dB dB
–50 dBm 10 dB 40 dB dB dBm dB 0 dB (reference)
–60 dBm 0 dB 50 dB dB dBm dB dB