2-5 Spectrum Analyzer Absolute Amplitude Accuracy Verification Spectrum Analyzer Verification
2-10 PN: 10580-00253 Rev. J S3xxE MM
11. Press the Amplitude soft key and then press the Reference Level soft key.
12. Use the keypad to enter 10 and press the dBm soft key.
13. Press the Atten Lvl soft key and enter 30, then press the dB soft key.
14. Adjust the source power so that the power meter displays the corresponding desired Sensor B reading as
recorded for 0 dBm in the Required Sensor B Reading column of Table A-5.
15. Press the Marker soft key and press the Peak Search soft key.
16. Record the Marker 1 amplitude reading in the 0dBm row of Table A-6, “Spectrum Analyzer 50 MHz
Absolute Amplitude Accuracy”.
17. Verify that the Marker 1 amplitude reading is within the specification.
18. Repeat Step 14 through Step 17 for the other power level settings. Refer to Table A-5 for Required
Sensor B Readings. Use Table A-6 to record test results. The last two power level settings are with the
pre-amp turned on, to ensure pre-amp functionality.
Amplitude Accuracy Across Frequency Verification
This procedure is the second test used to verify the absolute amplitude accuracy of the Spectrum Analyzer in
the S332E or S362E Site Master. The first procedure test was described above in “50 MHz Amplitude Accuracy
Verification” on page 2-7.
Equipment Required
• Anritsu MG3692X Synthesized Signal Source
• Anritsu ML2438A Dual Channel Power Meter
• Anritsu MA2442D High Accuracy Power Sensors (2)
• Anritsu 34NN50A 50 ohm Adapter
• Anritsu 34RKNF50 50 ohm Adapter
• Anritsu 15NN50-1.5C RF Coaxial Cable
• Aeroflex/Weinschel 1870A Power Splitter
• Aeroflex/Weinschel 44-10 10 dB Fixed Attenuator