7. SCPI Command Reference
43
Tx> TIME? //Or TIME:CHAR?
Rx> 12.0
Available under DISCH state
:SAMPle
The :SAMPle command sets sample timer value.
<float> fix float or floating point,0-999.9
Tx> TIME:SAMP 100.1
Tx> TIME:SAMP 0 //The sample timer sets to OFF
Available under DISCH state
7.12 APERture Subsystem
Figure 7-8
APERture
Command True
The APERture subsystem sets the convert time of the ADC.
APERture {slow,medium(med),fast}
7.13 SYSTem Subsystem
The SYSTem subsystem sets KEYLOCK to ON/OFF.