EasyManua.ls Logo

Applent Instruments AT682 - A Specifications

Default Icon
51 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A. Specifications
47
A Specifications
This chapter describes the specifications and supplemental performance characteristics
of the AT682/683
Specifications
Dimension
AT682/683 Megohmmeter Specifications
Accuracy is defined as meeting all of the following conditions.
Temperature: 23±5
Humidity: 65% R.H.
Zeroing: Open Correction
Warm up time is 30 min or more.
Rate: Slow
A 1-year calibration cycle
Insulation Resistance: Accuracy
range
voltage
1
2
3
4
5
6
1V
100k~1M
1M~10M
10M~100M
100M~1G
10V
100k~1M
1M~10M
10M~100M
100M~1G
1G~10G
25V
25k~250k
250k~2.5M
2.5M~25M
25M~250M
250M~2.5G
2.5G~25G
50V
50k~500k
500k~5M
5M~50M
50M~500M
500M~5G
5G~50G
75V
75k~750k
750k~7.5M
7.5M~75M
75M~750M
750M~7.5G
7.5G~75G
100V
100k~1M
1M~10M
10M~100M
100M~1G
1G~10G
10G~100G
125V
125k~1.25M
1.25M~12.5M
12.5M~125M
125M~1.25G
1.25G~12.5G
12.5G~125G
250V
250k~2.5M
2.5M~25M
25M~250M
250M~2.5G
2.5G~25G
25G~250G
500V
500k~5M
5M~50M
50M~500M
500M~5G
5G~50G
50G~500G
750V
750k~7.5M
7.5M~75M
75M~750M
750M~7.5G
7.5G~75G
75G~750G
1000V
1M~10M
10M~100M
100M~1G
1G~10G
10G~100G
100G~1T