Dimensions:
7KHUDSHXWLF6XUIDFHVLQÀDWHG
First Step Classic and Air Rail 82 in L x 35 in W x 5 in H / 208 cm L x 89 cm W x 12.7 cm H
First Step Cirrus 84 in L x 35 in W x 5 in H / 213 cm L x 89 cm W x 12.7 cm H
First Step Excel 82 in L x 35 in W x 7 in H / 208 cm L x 89 cm W x 18 cm H
Mattress Foam Base 82 in L x 35 in W x 3 in H / 208 cm L x 89 cm W x 7.6 cm H
V-Series Adaptor 84 in L x 35 in W x 5.5 in H / 213 cm L x 89 cm W x 14 cm H
MRS foam base is optional with all overlays.
Therapy Unit
115 VAC 12 in W x 10-3/4 in H x 7 in D / 30.5 cm L x 27.3 cm W x 18 cm H
Weight:
Therapeutic Surfaces:
First Step Classic and Air Rail 3 lb (1.3 kg)
First Step
Cirrus 7.5 lb (3.4 kg)
First Step Excel 4 lb (1.8 kg)
First Step Foam Base (standard) 5 lb (2.3 kg)
96HULHV$GDSWRU¿WV+LOO5RP9HUVD&DUH0HG6XUJ OENJ
Maximum Recommended Patient Weights
First Step Classic 250 lb (113 kg)
First Step
Cirrus 300 lb (136 kg)
First Step Air Rail 250 lb (113 kg)
First Step Excel 300 lb (136 kg)
Therapy Unit
115 VAC 14 lb (6.4 kg)
Electrical:
Voltage 115 VAC
)UHTXHQF\ +]
Amps 6 Amps
Max Electrical Leakage 100 Micro Amps
Fuse F1 - 1.0A, 250 VAC, Time Lag
F2 - 6.3A, 250 VAC, Time Lag
Fuse Breaking Capacity F1 - 50 Amps @ 115 VAC
F2 - 50 Amps @ 115 VAC
Power Cord Length 16 ft (4.9 m)