2.4.5 Sealing requirements for under-bench model
The following applies to the under-bench model FD1805.
If the outside of the machine is in contact with surrounding
surfaces, the machine must be waterproofed against these
surfaces to prevent water from leaking into the machine. This
applies to the machine’s top side (A), side panels (B) and back
side (C).
B
003242
C
B
A
A
Figure 8. Surfaces with sealing requirements
2.5 Water connections
2.5.1 General requirements
The pipes must be tted with an approved 1/2" shuto󰀨 valve.
Water connections should meet the requirements according to the
table below.
Cold water (Blue) Warm water (Red)
Connection 15 mm (1/2") 15 mm (1/2")
Pressure 100 - 800 kPa 100 - 800 kPa
Flow rate 20 l/min at 400 kPa
pressure
20 l/min at 400 kPa
pressure
Temperature At least 2°C, but less
than 20°C
45 – 60°C