Model E1070 Indicator Service Manual
AUDIT (Audit counters)
menu
The next section of the Supervisor menu is the AUDIT submenu. This menu
lets you view configuration and calibration audit counters. These counters
cannot be changed, only viewed.
Follow these steps to access each item in the AUDIT submenu:
1. From previous step ?, press the UNITS key. . .
AUDIT is displayed.
2. Press the PRINT key. . .
CFG is displayed. This stands for the Configuration audit counter.
Use this item to see how many times this indicator has been
configured.
3. Press the PRINT key. . .
A number is briefly displayed, then CFG is displayed. This is the
number of times this indicator has been configured.
4. Press the UNITS key. . .
CAL is displayed. This stands for the Calibration audit counter.
Use this item to see how many times this indicator has been
calibrated.
5. Press the PRINT key. . .
A number is briefly displayed, then CAL is displayed. This is the
number of times this indicator has been calibrated.
6. Press the SELECT key. . .
AUDIT is displayed.
This completes the AUDIT submenu and the Supervisor menu. Press the
SELECT key to return to normal weighing mode.
CFG
(Configuration audit counter)
CAL
(Calibration audit counter)
Supervisor Menu—(continued)