Model E1070 Indicator Service Manual
APP submenu
Figure 6
APP (applications) submenu
Follow these steps to access each item in the APP menu and to understand
what they do and how to set them:
1. Access the Service menu. . .
CAL is displayed.
2. Press the UNITS key twice. . .
APP is displayed.
3. Press the PRINT key. . .
SITE is displayed.
Use this item to choose your instrument location; NA (North
America), EU (Europe). Choosing the correct one will set defaults
to your location’s requirements.
4. Press the PRINT key. . .
Current setting is displayed.
5. Toggle between the choices by using the TARE or UNITS key and
press the ENTER key to accept the displayed choice. .
SITE is displayed.
See note at left.
SITE
(Setting site defaults)
The next section of the Service menu is the APP submenu. See Figure 6.
This menu lets you choose the default parameters for your location and also
lets you enable or disable each application available in this indicator. Under
each enabled application you can edit the default print format (#0) and
choose which formats (#0-10) to print and through which port. You can
configure the extra formats (#1-10) in the SERIAL submenu item in the
Service menu.
Applications are enabled in the
Service menu but you do each
application’s setup in the
Supervisor menu.
If you change the site setting
and save the change, then
change it back to the original
site and save, the defaults will
be reset to factory defaults.
This will not affect calibration,
print formats or recipes.
In normal operating mode,
press the MODE key to
change from one enabled
application to the next.
*
* Print format ‘0’ is used for this string. Strings
#1-10 are located under the Serial submenu.