Model E1070 Indicator Service Manual
Error # Error Description of Error Possible Cause
1 Communications error SensorComm not responding -Cable
-SensorComm hardware failure
-Indicator hardware failure
2 Power fault +Vin, +EXC, or -EXC has fallen out -Power supply failure
of tolerance. Voltage ±5%. -Cable
3 A to D overrange More than +5mV/V has been applied -Cable - weight sensor
to the A to D converter -Weight sensor failure
-SensorComm PC board
4 A to D underrange Less than -5mV/V has been applied -Cable - weight sensor
to the A to D converter -Weight sensor failure
-SensorComm PC board
5 A to D Initialization A to D converter not responding - SensorComm PC board
failure
6 Weight sensor overrange The weight sensor output has -Abuse of scale
exceeded the configured amount. -Weight sensor failure
7 Weight sensor deadload The weight sensor output has -Gauging problem on the weight
shift warning exceeded the configured amount sensor
of capacity since calibration. -Mechanical issue with the scale
8 Weight sensor deadload The output of the weight sensor has -Gauging problem on the weight
shift error increased more than a configurable sensor
percent of capacity since calibration -Mechanical issue with the scale
9 Weight sensor stability The output of 1 or more weight -Mechanical issue with the scale
sensor is not in the same range as -Weight sensor problem
the rest of the scale.
All messages below which mention components
are referring to components within the SensorComm product.
SensorComm Error Messages
Table 5
SensorComm errors