On-screen display (OSD)
58
5. Use Enter to confirm the selection.
You can select the following reading room types::
Reading room definition
This function is available only when the DICOM display function is
selected and Continuous DICOM ALC is off.
To modify a reading room definition:
1. Select the DICOM display function. See To select a display
function (DF):, on page 53.
Name Description
CR/DR/MAMMO Corresponds to light conditions in
diagnostic reading rooms for com-
puted radiology, digital radiology or
mammography. This setting has the
lowest maximum ambient light.
CT/MR/NM Corresponds to light conditions in
diagnostic reading rooms for com-
puted tomography, magnetic reso-
nance or nuclear medicine scans.
Office Corresponds to light conditions in
office rooms.
Clinical Corresponds to light conditions in
diagnostic reading rooms for clinical
viewing.
Emergency Corresponds to light conditions in
emergency rooms.
Operating Corresponds to light conditions in
operating rooms. This setting has
the highest maximum ambient
light.